EasyManua.ls Logo

Mindray BeneView T1 - Setting the Flow Rate; Setting up the CO Wave; Setting RR Source; Setting Barometric Pressure Compensation

Mindray BeneView T1
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15-7
15.4.7 Setting the Flow Rate
For the sidestream CO
2
module, you can change the sampling rate of respiratory gas in the patient’s airway by setting
the flow rate. To set the flow rate, enter the [CO2 Setup] menu and select an appropriate setting from [Flow Rate].
WARNING
Please consider the patient’s actual bearing capability and select the appropriate flow rate when setting the
flow rate.
15.4.8 Setting up the CO
2
Wave
In the [CO2 Setup] menu, you can:
Select [Wave Type] and toggle between [Draw] and [Fill]:
[Draw]: The CO
2
wave is displayed as a curved line.
[Fill]: The CO
2
wave is displayed as a filled area.
Select [Sweep] and then select the appropriate setting. The faster the wave sweeps, the wider the wave is.
Change the size of the CO
2
waveform by adjusting the wave [Scale].
15.4.9 Setting RR Source
To set RR source:
1. Enter the [CO2 Setup] menu.
2. Select [RR Source] and then select a source or [Auto] from the dropdown list.
The [RR Source] settings of Resp and CO
2
, are linked. For details, please refer to Setting RR Source of chapter Resp.
15.4.10 Setting Barometric Pressure Compensation
Both sidestream and microstream CO
2
modules have the function of automatic barometric pressure compensation (the
system automatically measures the barometric pressure which the patient monitor is exposed to). However, the
mainstream CO
2
module does not have such function. For the mainstream CO
2
module, the default barometric pressure
is 760 mmHg. You must modify the barometric pressure based on the actual situation as follows:
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the required password[Module
Maintenance >>][Maintain CO2 >>][Calibrate CO2 >>].
2. Select [Barometric Pressure] and then enter the value of barometric pressure to which the patient monitor is
exposed to.
WARNING
Be sure to set the barometric pressure properly before using the mainstream CO
2
module. Improper settings
will result in erroneous CO
2
reading.

Table of Contents

Related product manuals