C-1
C Default Configurations
This chapter lists some of the most important factory default settings in configuration management. You cannot change
the factory default configuration itself. However, you can make changes to the settings from the factory default
configuration and then save the changed configuration as a user configuration.
C.1 Parameters Configuration
C.1.1 ECG
ECG Setup
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Lead Set Yes Yes Auto (automatic lead type identification)
Alm Source Yes Yes HR
Alarm Yes Yes On
Alm Lev Yes Yes Med
HR/PR
High
Adu
Yes Yes
120
Ped 160
Neo 200
HR/PR
Low
Adu
Yes Yes
50
Ped 75
Neo 100
Sweep Yes Yes 25 mm/s
Beat Vol Yes Yes
General, OR: 2
ICU, NICU, CCU: 1
Paced No Yes No
Notch Filter Yes Yes Weak
Gain Yes Yes X1
Filter Yes Yes
General: Monitor
OR: Surgery
ICU, NICU: Monitor
CCU: Diagnostic
ECG Display Yes Yes Normal
Pacemaker Rate Yes Yes 60
Minimum QRS Threshold No Yes 0.16 mV