12-5
WARNING
Continuous non-invasive blood pressure measurements may cause purpura, ischemia and neuropathy in
the limb with the cuff. Inspect the application site regularly to ensure skin quality and inspect the extremity
of the cuffed limb for normal color, warmth and sensitivity. If any abnormity occurs, move the cuff to
another site or stop the blood pressure measurements immediately.
12.5.7 Sequence Measurement
NIBP sequence measurement can include up to five cycles: A, B, C, D and E. You can individually set the duration and
interval for each cycle.
To set the sequence measurement, follow this procedure:
1. Select the NIBP parameter window to enter the [NIBP Setup] menu.
2. Select [Sequence Setup>>]
3. Set up [Duration] and [Interval] for each cycle.
To start the sequence measurement, follow this procedure:
1. Select the NIBP parameter window to enter the [NIBP Setup] menu.
2. Set [Interval] to [Sequence]
3. Select [Start NIBP], or select [NIBP Measure] Quickkey in the main screen.