EasyManua.ls Logo

Mindray BeneView T1 - 23.9 Calibrating CO2

Mindray BeneView T1
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
23-7
23.9 Calibrating CO
2
For sidestream and microstream CO
2
modules, a calibration is needed every year or when the measured values have a
great deviation. For maintream CO
2
module, no calibration is needed. Calibration for sidestream CO
2
module can be
performed only when the sidestream module enters the full accuracy mode.
Tools required:
A steel gas cylinder with 6±0.05% CO2 and balance gas N2
T-shape connector
Tubing
Follow this procedure to perform a calibration:
1. Make sure that the sidestream or microstream CO
2
module has been warmed up or started up.
2. Check the airway for leakage and perform a leakage test as well to make sure the airway has no leakage.
3. Select [Main Menu] [Maintenance >>] [User Maintenance >>] enter the required password
[Maintain CO
2
>>] [Calibrate CO
2
>>].
4. In the [Calibrate CO
2
] menu, select [Zero].
5. After the zero calibration is finished successfully, connect the equipment as follows:
6. Turn on and adjust the relief valve to make the flowmeter reads within 10-50 mL/min and keeps stable as well.
7. In the [Calibrate CO
2
] menu, enter the vented CO
2
concentration in the [CO
2
] field.
8. In the [Calibrate CO
2
] menu, the measured CO
2
concentration is displayed. After the measured CO
2
concentration
becomes stable, select [Calibrate CO
2
] to calibrate the CO
2
module.
If the calibration is finished successfully, the message [Calibration Completed!] is displayed in the [Calibrate CO
2
] menu.
If the calibration failed, the message [Calibration Failed!] is displayed. In this case, perform another calibration.
Flowmeter
Tubing
T-shape connector
Monitor
Relief valve
Gas cylinder

Table of Contents

Related product manuals