13-1
13 Monitoring Temp
13.1 Making a Temp Measurement
This monitor can simultaneously monitor two temperature sites.
Verify that the probe detection program works correctly before monitoring. Plug out the temperature probe cable from
the T1 or T2 connector, and the monitor can display the message [T1 Sensor Off] or [T2 Sensor Off] and give alarm
tones correctly.
1. Select an appropriate probe for your patient according to patient type and measuring site.
2. If you are using a disposable probe, connect the probe to the temperature cable.
3 Plug the probe or temperature cable to the temperature connector.
4. Attach the probe to the patient correctly.
5. Check that the alarm settings are appropriate for this patient.
13.2 Understanding the Temp Display
The te
mperature monitoring is displayed on the monitor as three numerics: T1, T2 and TD. By selecting this area, you can
enter the [Alarm Setup] menu.
13.3 Changing Temperature Settings
13.3.1 Setting the Temperature Unit
Select [Unit Setup >>] from the [User Maintenance] menu. In the popup menu, select [Temp Unit] and toggle
between [ºC] and [ºF].
13.3.2 Setting the Temperature Label
The default temperature label is T1 and T2. To change the Temp label, follow this procedure:
1. Select the Temp parameter area to enter [Temp Setup] menu.
2. Select [Temp-1 Label] or [Temp-2 Label], and in the drop-down list, select a proper label.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T1 and is the answer not in the manual?

Mindray BeneView T1 Specifications

General IconGeneral
BrandMindray
ModelBeneView T1
CategoryMedical Equipment
LanguageEnglish

Related product manuals