EasyManua.ls Logo

Mindray BeneView T1 - 18.3 Oxygenation Calculations

Mindray BeneView T1
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18-3
18.2.3 Titration Table
To open the titration table, select [Titration Table >>] in the [Dose Calculation] window after the dose calculation is
finished.
In the titration table, when you change:
[Reference]
[Interval]
[Dose Type]
The titrated values change accordingly.
You can also:
Select
or , or or beside the vertical scrollbar to view more values.
Select [Record] to print out the currently displayed titrated values by the recorder.
18.2.4 Drug Calculation Formulas
Abbreviation Unit Formula
Conc. g/ml, unit/ml or mEq/ml Amount / Volume
Dose Dose/hr, Dose/kg/min Rate × Conc.
Volume ml Rate × Duration
Amount g, unit, mEq Rate × Duration
Duration h Amount/Dose
Drip Rate gtt/min INF Rate × Drop Size / 60
18.3 Oxygenation Calculations
18.3.1 Performing Calculations
To perform an oxygenation calculation:
1. Select [Main Menu][Calculations >>][Oxygenation >>], or select [Calculations] QuickKey
[Oxygenation >>].
2. Enter values for calculation.
3. Select the [Calculate] button. The system performs a calculation per the current settings and displays the
calculated values.
If a calculated value is outside the range, its background will highlight in yellow. You can select [Range] to
view its normal range in the unit field.
Invalid values are displayed as [---].

Table of Contents