16-10
16.8.2 Hemodynamic Parameters
Select [Hemodynamic Parameters] tab from the [Hemodynamic Parameters] menu to view the patient’s
hemodynamic parameters. In the [Hemodynamic Parameters] menu, you can select [Range] to view the referential
normal range of each parameter. If a parameter value exceeds its normal range, the system will add a ” or ” to the
right of the parameter.
Abbreviation Full Spelling Unit Default Normal Range
Output
CCO Continuous Cardiac Output L/min /
CCI Continuous Cardiac Index L/min/m
2
3.0-5.0
SV Stroke Volume ml /
SVI Stroke Volume Index ml/m
2
40-60
HR Heart Rate bpm 60-80
Contractility
GEF Global Ejection Fraction % 25-35
CFI Cardiac Function Index 1L/min 4.5-6.5
dPmx Left Ventricular Contractility mmHg/s /
Preload Volume
GEDV Global End Diastolic Volume ml /
GEDI Global End Diastolic Volume Index ml/m
2
680-800
ITBV Intrathoracic Blood Volume ml /
ITBI Intrathoracic Blood Volume Index ml/m
2
850-1000
SVV Stroke Volume Variation % 0-10
PPV Pulse Pressure Variation % 0-10
Afterload
SVR Systemic Vascular Resistance
DS/cm
5
or
kPa-s/l
/
SVRI Systemic Vascular Resistance Index
DS·m
2
/cm
5
or
kPa-s-m
2
/l
1700-2400
pArt-M Mean Artery Pressure
mmHg/kPa
or cmH
2
O
70-90
pArt-D Diastolic Artery Pressure
mmHg/kPa
or cmH
2
O
60-80
pArt-S Systolic Artery Pressure
mmHg/kPa
or cmH
2
O
100-140
Organ Function
EVLW Extravascular Lung Water ml /