EasyManua.ls Logo

Mindray BeneView T1 - 8.10 12-Lead ECG Monitoring

Mindray BeneView T1
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-22
8.10 12-Lead ECG Monitoring
8.10.1 Entering the 12-lead ECG Monitoring Screen
1. Refer to the section 8.3.3 ECG Lead Placements to place the electrodes.
2. In the [ECG Setup] menu, select [Others>>] to enter the [Others Setup Menu].
3. Set [Lead Set] to [12-Lead], set [ECG Display] to [12-Lead].
There are totally 12 ECG waves and 1 rhythm wave displayed on the screen. The rhythm lead is ECG I before entering the
12-lead ECG screen.
The [Filter] mode is automatically switched to [Diagnostic] when the patient monitor accesses the 12-lead full-screen;
the [Filter] mode resumes to the configuration before accessing the 12-lead full screen when the patient monitor exit
the 12-lead full screen.
8.10.2 Setting ECG Waveform Sequence
You can select the sequence of ECG waveforms on the 12-lead ECG screen and 12-lead ECG report.
To select the sequence of the ECG waveforms,
1. In the [ECG Setup] menu, select [Others>>] to enter the [Others Setup Menu].
2. Set [Waveform Layout] to [Standard] or [Cabrera].
[Standard]: the sequence is I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6.
[Cabrera]: the sequence is aVL, I, -aVR, II, aVF, III, V1, V2, V3, V4, V5, V6.
8.10.3 Extending the rhythm lead waveform area
You can extend the height of rhythm lead waveform area. To do so,
1. In the [ECG Setup] menu, select [Others>>] to enter the [Others Setup Menu].
2. Set [ECG Display Area] to [Extended].

Table of Contents

Related product manuals