23-2
23.2 Maintenance and Testing Schedule
The following maintenance and tests, except for visual inspection, power on test, touchscreen calibration, and battery
check, shall be carried out by the service personnel only. Contact your service personnel if any maintenance is required.
Make sure to clean and disinfect the equipment before any test and maintenance.
Check/Maintenance Item Recommended Frequency
Preventative Maintenance Tests
Visual inspection 1. When first installed or reinstalled.
NIBP test Pressure check 1. If the user suspects that the measurement is incorrect.
2. Following any repairs or replacement of relevant module.
3. At least once a year.
Leakage test
Sidestream and
Microstream CO2 tests
Leakage test
Performance test
Calibration
Performance Tests
ECG test and calibration Performance test 1. If the user suspects that the measurement is incorrect.
2. Following any repairs or replacement of relevant module.
3. At least once every two years.
Note: At least once a year is recommended for NIBP and CO2.
Calibration
Resp performance test /
SpO2 test /
NIBP test Pressure check
Leakage test
Temp test /
IBP test and calibration Performance test
Pressure calibration
Mainstream CO2 test and
calibration
/
Sidestream and
Microstream CO2 tests
and calibration
Leakage test
Performance test
Calibration
PiCCO test
Analog output
performance test
/ If the user suspects that the analog output does not work well.
Electrical Safety Tests
Electrical safety tests At least once every two years.
Other Tests
Power on test 1. When first installed or reinstalled.
2. Following any maintenance or the replacement of any main unit
parts.
Touchscreen calibration / 1. When the touchscreen appears abnormal.
2. After the touchscreen is replaced.
Battery check Functionality test 1. When first installed.
2. Whenever a battery is replaced.
Performance test Every two months or if the battery run time reduced significantly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T1 and is the answer not in the manual?

Mindray BeneView T1 Specifications

General IconGeneral
BrandMindray
ModelBeneView T1
CategoryMedical Equipment
LanguageEnglish

Related product manuals