EasyManuals Logo
Home>Mindray>Medical Equipment>BeneView T1

Mindray BeneView T1 User Manual

Mindray BeneView T1
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #158 background imageLoading...
Page #158 background image
15-6
15.4.4 Setting up Humidity Compensation
Sidestream and microstream CO
2
modules are configured to compensate CO
2
readings for either Body Temperature and
Pressure, Saturated Gas (BTPS), to account for humidity in the patient’s breath, or Ambient Temperature and Pressure,
Dry Gas (ATPD).
1. ATPD:
100/P%)vol(CO)mmHg(P
amb22co
2. BTPS:
100/)47P(%)vol(CO)mmHg(P
amb22CO
Where,
2CO
P
= partial pressure,
%vol
= CO
2
concentration,
amb
P
= ambient pressure, and unit is mmHg.
As the mainstream CO
2
module has a built-in heating component to prevent water vapour from condensing, setting
humidity compensation is not needed. For the sidestream and microstream CO
2
module, you can set the humidity
compensation on or off according to the actual condition. To set the humidity compensation:
1. In the [CO2 Setup] menu, select [BTPS Compen].
2. Select either [On] for BTPS or [Off] for ATPD, depending on which compensation applies.
15.4.5 Setting the Apnea Alarm Delay
In the [CO2 Setup] menu, select [Apnea Delay] and then select the appropriate setting. The monitor will alarm if the
patient has stopped breathing for longer than the preset apnea time. The [Apnea Delay] of Resp and CO
2
keeps
consistent with each other.
WARNING
The respiration measurement does not recognize the cause of apneas. It only indicates an alarm if no breath
is detected when a pre-adjusted time has elapsed since the last detected breath. Therefore, it cannot be
used for diagnostic purpose.
15.4.6 Choosing a Time Interval for Peak-Picking
For microstream and mainstream CO
2
modules, you can select a time interval for picking the highest CO
2
as the EtCO
2
and the lowest as the FiCO
2
.
In the [CO2 Setup] menu, select [Max Hold] and toggle between [Single Breath], [10 s], [20 s] and [30 s] (for
microstream CO
2
module only).
[Single Breath]: EtCO
2
and FiCO
2
are calculated for every breath.
[10 s], [20 s], or [30 s]: EtCO
2
and FiCO
2
are calculated using 10, 20 or 30 seconds of data.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T1 and is the answer not in the manual?

Mindray BeneView T1 Specifications

General IconGeneral
BrandMindray
ModelBeneView T1
CategoryMedical Equipment
LanguageEnglish

Related product manuals