EasyManuals Logo

Mindray BeneView T1 User Manual

Mindray BeneView T1
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #137 background imageLoading...
Page #137 background image
12-7
12.7.5 Switching On NIBP End Tone
The monitor can issue a reminder tone at the completion of NIBP measurement. The NIBP End Tone is off by default. You
can switch it on by accessing the [NIBP Setup] menu.
12.8 Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood vessel and therefore help venous
puncture.
1. Select [VeniPuncture >>] from the [NIBP Setup] menu. In the popup menu, verify that the [Cuff Press.] value is
appropriate. Change it if necessary.
2. Select [VeniPuncture].
3. Puncture vein and draw blood sample.
4. Select the on-screen QuickKey to deflate the cuff. The cuff deflates automatically after a set time if you do not
deflate it.
During measurement, the NIBP display shows the inflation pressure of the cuff and the remaining time in venous
puncture mode.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T1 and is the answer not in the manual?

Mindray BeneView T1 Specifications

General IconGeneral
BrandMindray
ModelBeneView T1
CategoryMedical Equipment
LanguageEnglish

Related product manuals