C-2
ST Analysis
Item Name
Configurable
Default
In Config Mode In Monitor Mode
ST Analysis Yes Yes General, OR, ICU, NICU: Off
CCU: On
Alarm Yes Yes Off
Alm Lev Yes Yes Med
ST-X High Yes Yes when ST Unit is mV:
when ST Unit is mm:
0.20
2.0
ST-X Low Yes Yes when ST Unit is mV:
when ST Unit is mm:
-0.20
-2.0
ISO Yes Yes -80 ms
J Yes Yes 48 ms
ST Yes Yes J + 60 ms
X represents I, II, III, aVR, aVL, aVF, or V1 to V6.
QT/QTc Analysis
Item Name
Configurable
Default
In Config Mode In Monitor Mode
QT Analysis Yes Yes Off
QTc Formula Yes Yes Hodges
Analysis Lead Yes Yes All
Arrh. Analysis
Arrhythmia Threshold Settings (Mindray)
Item Name
Configurable
Default
In Config Mode In Monitor Mode
PVCs High Yes Yes 10
Asys. Delay Yes Yes 5
Vtac Rate Yes Yes 130
Vtac PVCs Yes Yes 6
Multif. PVC’s Window Yes Yes 15
Tachy Yes Yes
Adu:
Ped:
120
160
Brady Yes Yes
Adu:
Ped:
50
75
Extreme Tachy Yes Yes
Adu:
Ped:
160
180
Extreme Brady Yes Yes
Adu:
Ped:
35
50
Vbrd Rate Yes Yes 40
Vbrd PVCs Yes Yes 5
Pause Time Yes Yes 2