C-3
Arrhythmia Threshold Settings (Mortara)
Item Name
Configurable
Default
In Config Mode In Monitor Mode
PVCs high Yes Yes 10
Asys. Delay Yes Yes 5
Vtac Rate Yes Yes 130
Vtac PVCs Yes Yes 6
Multif. PVC’s Window Yes Yes 15
Tachy Yes Yes Adu:
Ped:
120
160
Brady Yes Yes Adu:
Ped:
50
75
Arrhythmia Alarm Settings (Mindray)
Item Name
Configurable Default
In Config Mode In Monitor Mode Alarm Switch Alarm Level
Asystole Alarm Yes Yes On High
VFib/VTac Alarm Yes Yes On High
Vtac Alarm Yes Yes On High
Vent. Brady Alarm Yes Yes On High
Extreme Tachy Alarm Yes Yes On High
Extreme Brady Alarm Yes Yes On High
PVCs/min Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
R on T Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Run PVCs Alarm Yes Yes Off Low
Couplet Alarm Yes Yes Off Prompt
Multif. PVC Alarm Yes Yes Off Med
PVC Alarm Yes Yes Off Prompt
Bigeminy Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Trigeminy Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Tachy Alarm Yes Yes Off Med
Brady Alarm Yes Yes Off Med
PNP Alarm Yes Yes Off Prompt
PNC Alarm Yes Yes Off Prompt
Missed Beats Alarm Yes Yes Off Prompt
Nonsus. Vtac Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Vent. Rhythm Alarm Yes Yes Med
Pause Alarm Yes Yes Off Low
Irr. Rhythm Alarm Yes Yes Off Prompt
Afib Alarm Yes Yes Off Prompt