C-4
Arrhythmia Alarm Settings (Mortara)
Item Name
Configurable Default
In Config Mode In Monitor Mode Alarm Switch Alarm Level
Asystole Alarm Yes Yes On High
VFib Alarm Yes Yes On High
Vtac Alarm Yes Yes On High
PVCs/min Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
R on T Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Run PVCs Alarm Yes Yes Off Low
Couplet Alarm Yes Yes Off Prompt
Vent. Rhythm Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Bigeminy Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Trigeminy Alarm Yes Yes
General, OR, ICU, NICU: Off
CCU: On
Med
Tachy Alarm Yes Yes Off Med
Brady Alarm Yes Yes Off Med
PNP Alarm Yes Yes Off Prompt
PNC Alarm Yes Yes Off Prompt
Missed Beats Alarm Yes Yes Off Med
Multif. PVC Yes Yes Off Med
Irr. Rhythm Alarm Yes Yes Off Prompt
C.1.2 RESP
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alarm Yes Yes On
Alm Lev Yes Yes Med
RR High Yes Yes
Adu, Ped:
Neo:
30
100
RR Low Yes Yes
Adu, Ped:
Neo:
8
30
Apnea Delay Yes Yes
Adu, Ped:
Neo:
20
15
Lead Yes Yes II
Gain Yes Yes X2
Sweep Yes Yes 6.25 mm/s
Detection Mode Yes Yes Auto
RR Source No Yes Auto