8-25
Fridericia:
3
1
60
HeartRate
QTQTc
Framingham:
HeartRate
QTQTc
60
1154
Waveform Layout Standard, Cabrera Standard Select ECG lead sequence for displaying and
printing.
[Standard]: the sequence is I, II, III, aVR, aVL, aVF, V1,
V2, V3, V4, V5, V6.
[Cabrera]: the sequence is aVL, I, -aVR, II, aVF, III, V1,
V2, V3, V4, V5, V6.
Report setup
Menu item Option Default Description
Report Format
12×1, 6×2, 3×4+1 3×4+1
Selects the format of the 12-lead ECG report.
[12×1]: ECG waveforms are displayed in 12 lines.
[6×2]: ECG waveforms are displayed in 6 lines and 2
columns.
[3×4+1]: ECG waveforms are displayed in 3 lines
and 4 columns followed by the rhythm lead
waveform.
Median Complex On, Off Off Selects whether Median Complex is included on the
12-lead ECG report.
Median Complex displays a median complex
waveform for each lead and a rhythm lead
waveform of 10 seconds in 3x4+1 format. For each
waveform, short vertical lines are used to mark the
start of the P-wave and QRS complex, and the end
of the P-wave, QRS complex, and T-wave.
Measurements On, Off On Selects whether the measurement result is included
on the 12-lead ECG report.
Measurement result includes Vent. Rate, PR Interval,
QRS Duration, QT/QTc Interval, and P/QRS/T Axes.
Interpretation On, Off On Selects whether diagnoses are included on the
12-lead ECG report.
Interpretation
Summary
On, Off On Selects whether interpretation summary is included
on the 12-lead ECG report.
Note: If the [Interpretation] option is not enabled,
interpretation summary is not included on the
report even if [Interpretation Summary] is
enabled.