EasyManua.ls Logo

Mindray BeneView T1 - Page 231

Mindray BeneView T1
286 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
A-5
A.5.3 LEDs
Main unit
Alarm lamp 1 (two colors: yellow and red)
Power on LED 1 (green)
External power LED 1 (green)
Battery LED 1 (green)
T1 docking station
Connection status indicator 1 (green)
External power supply indicator 1 (green)
A.5.4 Audio Indicator
Speaker
Give alarm tones (45 to 85 dB), key tones, QRS tones; support PITCH TONE and
multi-level tone modulation; alarm tones comply with IEC60601-1-8.
A.5.5 Outputs
Analog Output
Standard
Meets the requirements of IEC60601-1 for short-circuit protection and leakage
current
ECG Analog Output
Bandwidth
(-3dB; reference frequency: 10Hz)
Diagnostic mode:
Monitor mode:
Surgical mode:
ST mode:
0.05 to 150 Hz
0.5 to 40 Hz
1 to 20 Hz
0.05 to 40 Hz
QRS delay
25 ms (in diagnostic mode, and with Paced off)
Sensitivity 1V/mV ±5%
PACE rejection/enhancement
Pace enhancement
Signal amplitude: Voh≥2.5V
Pulse width: 10ms±5%
Signal rising and falling time: ≤100µs
IBP Analog Output
Bandwidth
(-3dB; reference frequency:1Hz)
DC to 40 Hz
Max transmission delay 30 ms
Sensitivity 1 V/100 mmHg ±5%
Defib Sync Pulse
Output impedance
100Ω
Max time delay 35 ms (R-wave peak to leading edge of pulse)
Amplitude
High level: 3.5 to 5 V, providing a maximum of 10 mA output current;
Low level: < 0.5 V, receiving a maximum of 5 mA input current.
Pulse width 100 ms ±10%
Rising and falling time ≤1 ms

Table of Contents