Rise time (module PN: 115-027545-00)
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter neonatal sampling
line:
<330 ms@90 ml/min.
Measured with a DRYLINE II adult watertrap and a 2.5-meter adult sampling line:
<300 ms@120 ml/min
awRR measurement range 0 to 120 rpm
awRR measurement precision ±2 rpm
Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Note: The response time is the sum of the rise time and the delay time.
Effect of interference gases on CO
2
measurements
Gas Concentration (%) Quantitive effect*
N
2
O ≤60
±1 mmHg
Hal ≤4
Sev ≤5
Iso ≤5
Enf ≤5
Des ≤15 ±2 mmHg
*: means an extra error should be added in case of gas interference when CO
2
measurements are performed between 0 -
40mmHg.
Inaccuracy specifications are affected by the breath rate and I:E change. The end-tidal gas reading is within specification for
breath rate below 15BPM and I:E ratio smaller than 1:1 relative to the gas readings without breath.
Alarm limit Range Step
EtCO
2
High (mmHg) (low limit + 2) to 99
1 EtCO
2
Low (mmHg) 1 to (high limit - 2)
FiCO
2
High (mmHg) 1 to 99
awRR High (rpm)
Adult, pediatric:
Neonate:
(low limit + 2) to 100
(low limit + 2) to 150
1
awRR Low (rpm) 0 to (high limit - 2)
Microstream CO
2
Module
Standard Meet standard of ISO 21647
CO
2
Measurement range 0 to 99 mmHg
Accuracy*
0 to 38 mmHg:
39 to 99 mmHg:
±2 mmHg
±5% of the reading+0.08% of (the reading-38)
Accuracy drift Meet the requirement for measurement accuracy within 6 hours
* Accuracy applies for respiration rate up to 80 rpm. For respiration rate above 80 rpm, the accuracy is 4 mmHg or ±12% of
the reading, whichever is greater, for EtCO
2
exceeding 18 mmHg. For respiration rate above 60 rpm, the above accuracy can
be achieved by using the CapnoLine H Set for Infant/Neonatal. In the presence of interfering gases, the above accuracy is
maintained to within 4%.
Resolution 1 mmHg
Sample flow rate
7.5
15
50
+
ml/min
Initialization time 30 s (typical)