Loading...
C-8
PA Alarm Limit
PA-S High (mmHg) Yes Yes
Adu:
Ped, Neo:
35
60
PA-S Low (mmHg) Yes Yes
Adu:
Ped, Neo:
10
24
PA-M High (mmHg) Yes Yes
Adu:
Ped, Neo:
20
26
PA-M Low (mmHg) Yes Yes
Adu:
Ped, Neo:
0
12
PA-D High (mmHg) Yes Yes
Adu:
Ped, Neo:
16
4
PA-D Low (mmHg) Yes Yes
Adu:
Ped, Neo:
0
-4
CVP, LAP, RAP, ICP, UVP, P3-P4 Venous Pressure Alarm Limits
IBP-M High (mmHg) Yes Yes
Adu:
Ped, Neo:
10
4
IBP-M Low (mmHg) Yes Yes 0
CPP Alarm Limits
CPP High (mmHg) Yes Yes
Adu:
Ped:
Neo:
130
100
90
CPP Low (mmHg) Yes Yes
Adu:
Ped:
Neo:
50
40
30
Art, Ao, BAP, FAP, LV, P1-P2 Arterial Pressure Scale
Scale (mmHg) Yes Yes 0-160
PA Scale
Scale (mmHg) Yes Yes 0-30
CVP, LAP, RAP, ICP, UVP Scale
Scale (mmHg) Yes Yes 0-20
UAP, P3-P4 Venous Pressure Scale
Scale (mmHg) Yes Yes 0-80
IBP Overlapping Left Scale
Scale (mmHg) Yes Yes 0-160
IBP Overlapping Right Scale
Scale (mmHg) Yes Yes 0-20