C-14
Item Name
Configurable
Default
In Config Mode In Monitor Mode
High Alarm Interval (s) No Yes 10
Med Alarm Interval (s) No Yes 20
Low Alarm Interval (s) No Yes 20
Alarm Light on Alarm Reset No Yes On
Reset Other Bed's Alarms No Yes Off
Alarm Reset By Other Bed No Yes On
Minimum Alarm Volume Yes Yes General: 2
OR: 1
ICU, NICU, CCU: 2
Alarm Sound No Yes ISO
Reminder Tone No Yes On
Reminder Interval No Yes 3 min
ECGLeadOff Lev. No Yes Low
SpO
2
SensorOff Lev. No Yes Low
IBPSensorOff Lev. No Yes Med
Lethal Arrh. OFF No Yes Disable
Extended Arrh. No Yes Disable
Alarm Delay No Yes 6 s
ST Alarm Delay No Yes 30 s
Other Bed Disconnection Alm No Yes On
Wave Line No Yes Mediate
Outdoor Mode No Yes Manual
ECG Standard No Yes AHA
Notch Freq. No Yes 50 Hz
Data Transfer Method No Yes Off
Transferred Data Length No Yes 4 h
Para Switch Authority No Yes Unprotected
Parameter Switch Yes Yes
When [Para Switch Authority] is set
to [Protected]: Unselected
When [Para Switch Authority] is set
to [Unprotected]: Selected
SpO
2
Tone No Yes Mode 1