D-2
Measurement Alarm messages L Cause and solution
Missed Beats M*
Irr. Rhythm M*
Vent. Rhythm M*
Multif. PVC M*
Nonsus. Vtac M*
Pause L*
PNP M*
The pacer appears abnormal. Check the pacer.
PNC M*
Resp
Resp Apnea H
The respiration signal was so weak that the monitor cannot perform
respiration analysis. Check the patient’s condition and the Resp
connections.
Resp Artifact H
The patient’s heartbeat has interfered with his respiration. Check the
patient’s condition and the Resp connections.
SpO
2
SpO
2
Desat H
The SpO2 value has fallen below the desaturation alarm limit. Check
the patient’s condition and check if the alarm limit settings are
correct.
No Pulse H
The pulse signal was so weak that the monitor cannot perform pulse
analysis. Check the patient’s condition, SpO
2
sensor and measurement
site.
CO
2
CO
2
Apnea H
The patient stops breathing, or the respiration signal was so weak
that the monitor cannot perform respiration analysis. Check the
patient’s condition and the RM connections.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneView T1 and is the answer not in the manual?

Mindray BeneView T1 Specifications

General IconGeneral
BrandMindray
ModelBeneView T1
CategoryMedical Equipment
LanguageEnglish

Related product manuals