Batteries 14-1
14 Batteries
WARNING:
1.
The battery is inside the machine; only technical
professionals from Mindray or engineers
authorized by Mindray after training can perform
battery installation and uninstallation.
2.
If you need to change the battery or buy a new
one, please contact your sales representative.
14.1 Overview
The battery charges when the system is connected to the AC power supply.
z Generally, a completely discharged battery will fully charge in less than 8 hours.
z When the battery capacity is charged to the full capacity, it can support a
standby period of no less than 24 hours.
Note: 1. Charge the battery in time when the battery capacity is empty is necessary;
otherwise, the performance of the batteries is degraded.
2. Power off the system if you will not use the system for a long period of time
(including storage/ transportation condition), and you should not allow the
system in standby status, otherwise the batteries will be out of power and
permanently damaged.
When the external power supply is not connected, two lithium-ion batteries will provide
the power.
Fully charged battery can ensure a normal working condition of more than 1 hour (B
mode with probe 3C5A probe).
CAUTION:
Only move the machine in the power off or standby status or
damage may happen.
14.2 Battery Status Indicator
The battery status indicator is located in the lower right corner of the screen, indicating
the battery capacity.
z : Indicates the battery capacity is empty.
z : Indicates the battery capacity is full.
z The battery icons other than the two icons above-mentioned indicate the
capacity is between empty and full.
When the power capacity is not enough, the system will pop up a prompt:
“Warning! Battery is out of power! Please connect to AC power or system will be shut
down in one minute.” Connect to AC power to afford normal work.