2-2 System Overview
2.4.2 Power supply
Voltage
100-240V~
Frequency 50/60Hz
Power
consumption
600VA
Fuse 250V~ T10AH
2.4.3 Environmental Conditions
Operational Conditions Storage and Transportation
Conditions
Ambient
temperature
0℃~40℃
10℃~40℃ (for D6-2 and D6-2A)
-20℃~55℃
-10℃~60℃ (for D6-2 and D6-2A)
Relative
humidity
30%~85% (no condensation)
25%~90% (for D6-2 and D6-2A)
30%~95% (no cage ondensation)
Atmospheric
pressure
700hPa~1060hPa 700hPa~1060hPa
2.4.4 Size and Weights
External dimensions:
z Highest: 1415mm (H)×500mm (W)×690mm (D)
z Lowest: 1275mm (H)×500mm (W)×740mm (D)
Net weight: 62Kg (without batteries and 4D board)
2.5 System Configuration
2.5.1 Standard Configuration
Main unit
Accessories
z Operator's manuals
z Ultrasound gel
z Ultrasound gel holder
z System power line
z Dust-proof cover
z Probe ports dust-proof cover
z 4D probe port dust-proof cover
z Grounding line
z Video printer remote line
z Hard disk (1T)
WARNING:
Do not use this system in the conditions other than those
specified.