System Maintenance 17-5
Content Description
Graph / text
printer
First wipe off dust or stain attached to the cover of printer with soft dry
cloth, then clean the inside of printer. Be sure to do the cleaning
maintenance according to the operation manual if is necessary.
Foot switch
Use soft dry cloth with a little mild soap water to wipe off the dust or stain
attached to the pedals or cable of foot switch.
Bar code
scanner
First use soft dry cloth to wipe off dust attached to glass panel of
scanner, then the dust or strain attached to cable and bracket. Be sure to
do the especial cleaning maintenance according to the operation manual
if is necessary.
17.1.3 Checking the Probe
îš„ Visually check to confirm that there is no crack and expansion to probe head.
îš„ Visually check to confirm that there is no deterioration or desquamation to probe
cable.
îš„ Visually check to confirm that there is no bend, destroyed or falling off pins to the
connector.
17.1.4 Checking the Power Cable and Plug
îš„ Visually check to confirm that there is no wrinkles, crack or deterioration; No crack
and expansion to the surface of adapter.
îš„ Manually check to confirm that there is no looseness or rupture. The connection of
plug is reliable.
17.1.5 Checking the Circuit Breaker
Manually check to confirm that the circuit breaker can be turned on or off normally.
17.1.6 Checking Appearance
Check if there is any crack in the covers:
îš„ Ultrasound system covers
îš„ Probe appearance
17.1.7 Backup of the System Hard Drive
To prevent deterioration or loss of data stored in the system hard drive, create a backup
copy of the hard drive at regular intervals.