Probes and Biopsy 13-13
z Relative humidity: 30% ~ 95% (no condensation)
z Atmospheric pressure: 700 hPa ~ 1060 hPa
Store and transport the D6-2 probe under the following ambient conditions:
z Ambient temperature: -10°C ~ 60°C
z Relative humidity: 30% ~ 95% (no condensation)
z Atmospheric pressure:700 hPa ~ 1060 hPa
Store and transport the D6-2A probe under the following ambient conditions:
z Ambient temperature: -10°C ~ 60°C
z Relative humidity: 30% ~ 90% (no condensation)
z Atmospheric pressure:700 hPa ~ 1060 hPa
3. When the probe is sent to MINDRAY Customer Service Department or sales
representative for repair, be sure to disinfect it and keep it in the carrying case to
prevent infection.
4. Sterilize the carrying case as necessary.
13.2 Biopsy Guide
WARNING:
1. The person performing biopsy procedures must
understand diagnostic ultrasound thoroughly and have
been trained adequately, otherwise, side effects may be
caused to the patient.
2. In situations listed below, the biopsy needle may fail to
penetrate the target. The incorrect biopsy may cause
various side effects in the patient.
Use a needle-guided bracket other than that provided.
Mount the needle-guided bracket incorrectly.
Use a biopsy needle that is unsuitable for the type of
biopsy being performed.
Use a biopsy needle or guiding block that is unsuitable
for the needle guide.
3. Before and after a biopsy procedure is performed,
confirm that the needle-guided bracket is normal.
Manually confirm that the parts of the needle-guided
bracket do not slip off or move from their proper
positions. If the needle-guided bracket is used when
parts are not securely and correctly installed, the
patient may be injured. If an abnormality is found on
the needle-guided bracket, immediately stop using it
and contact MINDRAY Customer Service Department or
sales representative.
4. DO NOT use a needle-guided bracket when scanning is
performed. The needle may advance in an incorrect
direction and possibly injure the patient.
Never perform a biopsy during image scanning.
5. DO NOT freeze an image while performing biopsy
procedure.