5-6 Image Optimization
Impacts The function is available in real-time imaging, freeze or cine review status.
iBeam
Description This function is used to superimpose and average images of different
steer angles to obtain image optimization.
Operation Adjust through the [iBeam] item in the soft menu.
Effects Images after iBeam processing can be optimized with less spot noise and
higher resolution, so that more details for the structure are revealed.
Impacts When linear probe is being used, iBeam function is not available when
trapezoid function is turned on or B steer is adopted.
Auto Merge
Description In the Dual-split mode, when the images of the two windows use the same
probe type, depth, invert status, rotation status and magnification factor,
the system will merge the two images so as to extend the field of vision.
Operation Turn on or off the function through the [Auto Merge] item in the soft menu.
Impacts iBeam is available only for linear probes.
The function is available in real-time imaging, freeze or cine review status.
Gray Map
Description This function applies the gray correction to obtain the optimum images.
Operation Select among the maps through the [Gray Map] item in the soft menu.
The system provides 8 groups of gray map.
Impacts The function is available in real-time imaging, freeze or cine review status.
Tint Map
Description Colorize function provides an imaging process based on color difference
rather than gray distinction.
Operation Select the colorize map through the [Tint Map] item in the soft menu or
menu.
The system provides 16 (1-16) Tint Maps to be selected.
Impacts The function is available in real-time imaging, freeze or cine review status.
TSI (Tissue Specific Imaging)
Description The TSI function is used to optimize the image by selecting acoustic
speed according to tissue characteristics.
Operation Select among the TSI modes through the [TSI] item in the soft menu.
The system provided 4 ways of optimization for specific tissues: general,
muscle, fluid and fat.
iTouch
Description To optimize image parameters as per the current tissue characteristics for
a better image effect.