Image Optimization 5-9
Speed
Description This function is used to set the scanning speed of M mode imaging, and
the real-time speed value is displayed in the image parameter area in the
upper left corner of the screen.
Operation Change the speed through the [Speed] item in the soft menu.
Or adjust it in the image parameter area.
There are 6 levels of scan speed available, the smaller the value the faster
the speed.
Effects Speed changing makes it easier to identify disorders in cardiac cycles.
Tint Map
Description Colorize function provides an imaging process based on color difference
rather than gray distinction.
Operation Select the colorize map through the [Tint Map] item in the soft menu.
The system provides 16 different maps to be selected among.
Impacts The function is available in real-time imaging, freeze or cine review status.
Gray Map
Description This function applies the gray correction to obtain the optimum image
through adjusting 3 parameters: curve, gray rejection and γ.
Operation Select among the maps through the [Gray Map] item in the soft menu.
You can also adjust through the grayscale bar: move the cursor to the
grayscale bar and press <Set> on the control panel to adjust.
The system provides 8 groups of gray map.
Impacts The function is available in real-time imaging, freeze or cine review status.
Edge Enhance
Description This function is used to increase image profile, so as to distinguish the
image boundary.
Operation Adjust through the [Edge Enhance] item in the soft menu.
The system provides 14 levels of edge enhance effects, off represents no
edge enhance is turned on, and the bigger the value the stronger the
effect.
Impacts Larger edge enhance may lead to noise increasing.
Dynamic Range
Description This function is used to adjust the B image resolution to compress or
expand the gray display range. The real-time dynamic range value will be
displayed in the image parameter area in the upper left corner of the
screen.
Operation Adjust through the [Dyn Ra.] item in the soft menu.
Or adjust it in the image parameter area.
The adjusting range is 30-240dB in increments of 5.