EasyManua.ls Logo

Mindray uMEC12 - Deleting a Reference Segment; Recording the ST Segment; Changing the ST Alarm Limits; Setting the ST Alarm Delay Time

Mindray uMEC12
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-14
8.6.6 Deleting a Reference Segment
To delete the current ST reference segment, select [Delete Ref.] in the [ST Analysis] menu
and then select [Ok] in the popup.
8.6.7 Recording the ST Segment
To record the current ST segment and reference segment, select [Record] in the [ST Analysis]
menu.
8.6.8 Changing the ST Alarm Limits
High and low ST alarm limits can be set individually for each ECG lead. Alarm limits can
also be set separately for single-lead and multi-lead ST monitoring. You can select [ST
Alarm Setup >>] from [ST Analysis] menu and then change ST alarm settings for each lead.
8.6.9 Setting the ST Alarm Delay Time
To set the ST alarm delay time,
1. Select [Main Menu][Maintenance >>][User Maintenance >>]. Enter the
required password and then select [OK].
2. Select [Alarm Setup >>][ST Alarm Delay].
8.6.10 Adjusting ST Measurement Points
As shown in the figure below, the ST measured for each beat complex is the vertical
difference between two measurement points with the R-wave peak as the baseline for the
measurement.
R-wave peak
P
T
Isoelectric point ST measurement point
Difference=ST value
J point
Q S

Table of Contents

Other manuals for Mindray uMEC12

Related product manuals