5-6
5.9 Restoring the Latest Configuration Automatically
During operation, you may make changes to some settings. However, these changes may not
be saved as user configuration. To prevent the changes from losing in case of a sudden power
failure, the patient monitor stores the configuration in real time. The saved configuration is
the latest configuration.
The monitor restore the latest configuration if restarts within 60 seconds after the power
failure. And it will restore the default configuration rather than the latest configuration if
restarts 120 seconds later after the power failure. The monitor may load either the latest
configuration or the default configuration if restarts from 60-120 seconds after the power
failure.
5.10 Modifying Password
To modify the password for accessing the [Manage Configuration] menu,
1. Select [Modify Password >>] in the [Manage Configuration] menu.
2. Input a new password in the popup menu.
3. Select [Ok].

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz