5-6 
 
5.9 Restoring the Latest Configuration Automatically 
During operation, you may make changes to some settings. However, these changes may not 
be saved as user configuration. To prevent the changes from losing in case of a sudden power 
failure, the patient monitor stores the configuration in real time. The saved configuration is 
the latest configuration. 
 
The monitor restore the latest configuration if restarts within 60 seconds after the power 
failure. And it will restore the default configuration rather than the latest configuration if 
restarts 120 seconds later after the power failure. The monitor may load either the latest 
configuration or the default configuration if restarts from 60-120 seconds after the power 
failure. 
 
 
5.10 Modifying Password 
To modify the password for accessing the [Manage Configuration] menu, 
1. Select [Modify Password >>] in the [Manage Configuration] menu. 
2.  Input a new password in the popup menu. 
3. Select [Ok].