16-5
CAUTION
z To avoid blocking the airway, empty the DRYLINE II watertrap container
whenever half full. Replacing the DRYLINE II watertrap once a month is
recommended.
z The DRYLINE PRIME watertrap mini is automatically blocked when its service
life is reached. Replace the DRYLINE PRIME watertrap mini if the alarm “CO2:
Change Watertrap” occurs.
z The watertrap has a filter preventing bacterium, water and secretions from
entering the module. Extended use could destroy the filter in watertrap and fail to
stop the bacterium, water and secretions entering the module, result in damaging
the gas module and having infection risk..
NOTE
z Do not apply adult watertrap to the neonate patient. Otherwise, patient injury
could result.
z To extend the lifetime of the watertrap and module, disconnect the watertrap from
the module and set the operating mode to Standby mode when CO
2
monitoring is
not required.
z The sapmle rates are different when different types of watertraps are used.
16.5.2 Zeroing the CO
2
Module
The CO
2
module performs zero calibration automatically when needed.
NOTE
z The CO
2
module temporally stops measuring during zeroing.

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals