7-6
 Mode 2:
 Interval between high level alarm tones: 1 s.
 Interval between medium level alarm tones: 5 s.
 Interval between low level alarm tones: 20 s.
If you choose the ISO pattern, you can change the interval between alarm tones. To change
the interval between alarm tones:
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [High Alarm Interval (s)], [Med Alarm Interval (s)] and [Low Alarm
Interval (s)] in turn and then select the appropriate settings.
WARNING
z When the alarm sound is switched off, the patient monitor will give no audible
alarm tones even if a new alarm occurs. Therefore the user should be very careful
about whether to switch off the alarm sound or not.
z Do not rely exclusively on the audible alarm system for patient monitoring.
Adjustment of alarm volume to a low level may result in a hazard to the patient.
Always keep the patient under close surveillance.
7.4.4 Changing the Alarm Tone Pattern
To change the alarm tone pattern:
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Alarm Sound] and toggle between [ISO], [Mode 1]and [Mode 2].
User or factory default configurations exert no impact on the setup of alarm tone pattern.
The alarm tone pattern remains unchanged after the monitor restarts.

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals