7-15
7.8 Resetting Alarms
By selecting the QuickKey, you can reset the alarm system to acknowledging the
on-going alarms and enable the alarm system to respond to a subsequent alarm condition.
For physiological alarms, except the NIBP-related alarms, when the alarm system is reset:
 The alarm sound is reset.
 A Ĝ appears before the alarm message, indicating that the alarm is acknowledged.
 The icon
appears in the alarm symbol area.
 The parameter numeric and alarm limits still flash.
The indication of alarm lamp for the physiological alarm depends on the alarm light setting.
 When [Alarm Light on Alarm Reset] is set to [On], the alarm lamp remains flashing.
 When [Alarm Light on Alarm Reset] is set to [Off], the alarm lamp stops flashing.
Technical alarms give different alarm indicators when the alarm system is reset:
 For some technical alarms, including the NIBP-related alarms, a Ĝ appears before
the alarm message and
appears in the alarm symbol area, indicating that the alarm
is acknowledged.
 Some technical alarms are changed to the prompt messages.
 Some technical alarms are cleared. The monitor gives no alarm indications.
For details about the indications of technical alarms when the alarm system is reset, refer to
D.2 Technical Alarm Messages.
To set [Alarm Light on Alarm Reset]:
1. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the
required password.
2. Select [Alarm Setup >>] to enter the [Alarm Setup] menu.
3. Select [Alarm Light on Alarm Reset], and toggle between [On] and [Off]. The
default setting for [Alarm Light on Alarm Reset] is [On].

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals