4-1 
4 Managing Patients 
 
 
4.1 Discharging a Patient 
Before monitoring a new patient, discharge the previous patient. After the patient is 
discharged, all patient data, including patient information, trend data, and physiological alarm 
information is deleted from the monitor. The technical alarms are reset, and monitor settings 
return to their defaults. For details, refer to 5.3 Setting Default Configuration.  
 
After a patient is discharged, the monitor automatically admits a new patient. 
 
 WARNING 
z  Always discharge the previous patient before starting monitoring a new patient. 
Failure to do so can lead to data being attributed to the wrong patient. 
 
 
NOTE
z  Discharging a patient clears the current patient data from the monitor.
 
 
4.1.1 Auto Discharging a Patient after Monitor Power Off 
You can let the monitor automatically discharge after the monitor has been switched off for a 
period of time. 
 
To set the time period of discharging a patient:   
1.  Select [Main Menu]ė[Maintenance >>]ė[User Maintenance >>]ėenter the 
required passwordė[OK]. 
2.  Select a time for [Auto Discharge Pat.]. The monitor will automatically discharge the 
patient when the selected time period is reached. The default is Never. That is to say the 
monitor will not discharge a patient no matter for how long the monitor has been 
switched off.