E.4 Protective Earth Resistance   
1.  Plug the probes of the analyzer into the device’s protective earth terminal and protective 
earth terminal of the AC power cord.   
2.  Test the earth resistance with a current of 25 A. 
3.  Verify the resistance is less than limits. 
 
LIMITS 
For all countries, R = 0.2  Maximum   
 
 
E.5 Earth Leakage Test 
Run an Earth Leakage test on the device being tested before performing any other leakage 
tests. 
 
The following outlet conditions apply when performing the Earth Leakage test:   
  normal polarity (Normal Condition) 
  reverse polarity (Normal Condition) 
  normal polarity with open neutral (Single Fault Condition) 
  reverse polarity with open neutral (Single Fault Condition) 
 
LIMITS 
For UL60601-1, 
  300 A in Normal Condition 
  1000 A in Single Fault Condition 
For IEC60601-1, 
  500 A in Normal Condition 
  1000 A in Single Fault Condition