E-3
E.4 Protective Earth Resistance
1. Plug the probes of the analyzer into the device’s protective earth terminal and protective
earth terminal of the AC power cord.
2. Test the earth resistance with a current of 25 A.
3. Verify the resistance is less than limits.
LIMITS
For all countries, R = 0.2  Maximum
E.5 Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other leakage
tests.
The following outlet conditions apply when performing the Earth Leakage test:
 normal polarity (Normal Condition)
 reverse polarity (Normal Condition)
 normal polarity with open neutral (Single Fault Condition)
 reverse polarity with open neutral (Single Fault Condition)
LIMITS
For UL60601-1,
 300 A in Normal Condition
 1000 A in Single Fault Condition
For IEC60601-1,
 500 A in Normal Condition
 1000 A in Single Fault Condition

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals