7-14
You can also temporarily prolong the alarm pause time after the monitor entersthe alarm
paused status:
1. Select [Main Menu]Ä—[Maintenance >>]Ä—[User Maintenance >>]Ä—enter the
required passwordÄ—[Alarm Setup >>].
2. In the [Alarm Setup] menu, set the [Max. Alarm Pause 15min] to [Enable].
3. In the physiological alarm area, select a proper time in the [Alarm Pause Time] menu.
7.7 Switching Off All Alarms
If [Alarm Pause Time] is set to [Permanent], the patient monitor enters into the alarm off
status after the
hardkey is pressed. During the alarm off status,
 As for physiological alarms, no alarm lamps flash and no alarms are sounded.
 As for physiological alarms, no numeric and alarm limit flash.
 No physiological alarm messages are shown.
 [Alarm Off] is displayed in the physiological alarm area with red background.
 As for technical alarms, no alarms are sounded.
 The
alarm off symbol is displayed in the alarm symbol area.
You can cancel the alarm off status by pressing the
hardkey.
WARNING
z Pausing or switching off alarms may result in a hazard to the patient. Please be
very careful.
NOTE
z [Max. Alarm Pause 15min] is configured to [Disable] by default. In this case, you
cannot prolong the pause time.The prolonged pause time is only effective to the
current paused alarms.