EasyManuals Logo

Mindray uMEC12 Operator's Manual

Mindray uMEC12
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #138 background imageLoading...
Page #138 background image
12-6
The STAT mode initiates 5 minutes of continuous, sequential, automatic NIBP measurements.
WARNING
z Continuous non-invasive blood pressure measurements may cause purpura,
ischemia and neuropathy in the limb with the cuff. Inspect the application site
regularly to ensure skin quality and inspect the extremity of the cuffed limb for
normal color, warmth and sensitivity. If any abnormity occurs, move the cuff to
another site or stop the blood pressure measurements immediately.
12.6 Understanding the NIBP Numerics
The NIBP display shows numerics only as below. Your display may be configured to look
slightly different.
(1) Time of last measurement
(2) Time remaining to next measurement
(3) Measurement mode
(4) Unit of pressure: mmHg or kPa
(5) Systolic pressure
(6) Diastolic pressure
(7) Mean pressure obtained after the measurement and cuff pressure obtained during the
measurement
If the NIBP measurement exceeds the measurement range, “---” will be displayed.
(3)
(4)
(5)
(6)
(7)
(1) (2)

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals