16-2
16.3 Measurement Limitations
The following factors may influence the measurement accuracy:
 Leaks or internal venting of sampled gas
 Mechanical shock
 Cyclic pressure up to 10 kPa (100 cmH
2
O)
 Other sources of interference, if any
Measurement accuracy may be affected by the breath rate and inspiration/expiration (I/E)
ratio as follow:
 EtCO
2
value is within specification for breath rate  60 bpm and I/E ratio  1: 1;
 EtCO
2
value is within specification for breath rate  30 bpm and I/E ratio  2: 1.
Measurement accuracy is unspecified for breath rate larger than 60 bpm.
16.4 Understanding CO
2
Display
The CO2 parameter and waveform areas provide FiCO2 measurement, EtCO2 measurement,
awRR measurement, and a CO
2 waveform.
(1) CO
2 waveform
(2) Fraction of inspired CO
2
(FiCO2)
(3) End tidal CO
2
value (EtCO2)
(4) Airway respiration rate (awRR)
(1)
(3)
(4)
(2)

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals