4-2
4.1.2 Manually Discharging a Patient
To manually discharge a patient:
1. Select the [Patient Setup] QuickKey, or [Main Menu][Patient Setup >>].
2. Select [Discharge Patient]. In the popup menu, you can either:
 Directly select [Ok] to discharge the current patient, or
 Select [Standby] then [Ok]. The patient monitor enters the standby mode after
discharging the current patient, or
 Select [Cancel] to exit without discharging the patient.
4.2 Admitting a Patient
4.2.1 Auto Admitting a Patient
The monitor automatically admits a new patient in the following situations:
 After a patient is manually discharged, the monitor automatically admits a new patient.
 After being switched off for the selected time period, the monitor automatically
discharges the previous patient and admits a new patient at startup.
Always inputs patient information as soon as the patient is admitted. For details, refer to 4.3.1
Editing Patient Information.
WARNING
z The settings of patient category and paced status always contain a default value,
regardless of whether the patient is admitted or not. Check if the setting is correct
for your patient.
z For paced patients, you must set [Paced] to [Yes]. If it is incorrectly set to [No], the
patient monitor could mistake a pace pulse for a QRS and fail to alarm when the
ECG signal is too weak.
z For non-paced patients, you must set [Paced] to [No].