7-17 
 
7.10 Testing Alarms 
When the monitor starts up, a selftest is performed. In this case the alarm lamp is lit in 
yellow and red respectively, and the system gives a beep. This indicates that the visible and 
audible alarm indicators are functioning correctly. 
 
For further testing of individual measurement alarms, perform the measurement on yourself 
(for example SpO
2
 or CO
2
) or use a simulator. Adjust alarm limits and check that 
appropriate alarm behaviour is observed. 
 
 
7.11 Using Care Group Alarms 
7.11.1 Care Group Auto Alarms 
When a Care Group is set up on your monitor, a flashing symbol will appear beside the 
QuickKeys area if any monitor in your Care Group, which is not currently viewed by your 
monitor, is alarming. The alarm symbol is shown as below. 
 
The background colors of the alarm symbols indicate alarm levels, and are the same as 
those of the corresponding alarm messages. If multiple alarms are active in the Care Group, 
the background color is the same as that of the highest-level alarm message. For more 
information about the alarm message and background color, see7.3.2 Alarm Message.  
 
When a patient monitor in the Care Group is disconnected, the flashing symbol is shown as 
below. 
 
The department and bed label of the alarming monitor appear on the symbol. You can enter 
the view other patient window by pressing this symbol. 
 
 
7.11.2 Setting Care Group Alert Tone   
When a monitor in the Care Group issues an alarm, your patient monitor prompts you by 
giving an alert tone. To set the alert tone:   
1.  In the main menu, select [Screen Setup >>]ė[Screen Layout >>]ė[Choose 
Screen]. 
2. In the [Choose Screen] window, select [View Others Screen]. 
3.  In the view other patient window, select [Setup>>], and set [Alert Tone] to [Repeat], 
[Once] or [Off].