15-2
15.3 Influencing Factors
The factors that affect cardiac output are:
 temperature of injectate solution,
 volume of injectate solution,
 patient’s baseline blood temperature,
 patient’s inspiratory/expiratory cycle,
 placement of catheter with relation to proximity of lung field,
 the catheter itself,
 the patient rhythm and hemodynamic status, and
 any other rapid IV solutions which are infused while the C.O. measurement is being
performed.
Followings are some technique suggestions to obtain accurate C.O.:
 Injectate solution must be cooler than the patient’s blood.
 Inject solution rapidly and smoothly.
 Inject at end expiration.
15.4 Setting Up the C.O. Measurement
WARNING
z Use only accessories specified in this manual. Make sure that the accessories never
come into contact with conductive parts.
1. Connect the C.O. cable to the C.O. connector on the monitor.
2. Interconnect the patient monitor, catheter and syringe as shown below. Make sure that:
 The PA catheter is in place in the patient.
 The C.O. cable is properly connected to the monitor.

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals