14-6
NOTE
z When two pressures are deteted having the same label, the patient monitor
changes one pressure label to a currently unused one.
14.6.2 Setting the Pressure Label Order
Select [IBP Label Order Setup >>] from the parameter setup menu to set the display order
of the pressure labels. The default display order is: Art, pArt, CVP, pCVP, ICP, PA, Ao, UAP,
FAP, BAP, LV, LAP, RAP, UVP, P1, P2, P3, P4. To restore the default setting, you can select
[Defaults] from the [IBP Label Order Setup] window
14.6.3 Setting Alarm Properties
Select [Alarm Setup >>] from the parameter setup menu. You can set alarm properties for
this parameter in the popup menu.
14.6.4 Changing Averaging Time
The IBP value displayed on the monitor screen is the average of data collected within a
specific time. The shorter the averaging time is, the quicker the patient monitor responds to
changes in the patient’s blood pressure. Contrarily, the longer the averaging time is, the
slower the patient monitor responds to changes in the patient’s blood pressure, but the
measurement accuracy will be improved. For critically ill patients, selecting shorter
averaging time will help understanding the patient’s state.
To set the averaging time, in the parameter setup menu, select [Sensitivity] and toggle
between [High], [Med] and [Low], the corresponding averaging time is about 1 s, 8 s and 12
s respectively.
14.6.5 Setting Up the IBP Wave
In the setup menu for the pressure, you can:
 Select [Sweep] and then select the appropriate setting. The faster the wave sweeps, the
wider the wave is.
 Select [Scale] and then select the appropriate setting. If [Auto] is selected, the size of
the pressure’s waveform will be adjusted automatically.
 Select [Filter] and then select the desired option.

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz