25-2
25.2 Maintenance and Testing Schedule
The following maintenance and tests, except for visual inspection, power on test, touchscreen
calibration, battery check, and recorder check, shall be carried out by the service personnel
only. Contact your service personnel if any maintenance is required. Make sure to clean and
disinfect the equipment before any test and maintenance.
Check/Maintenance Item Recommended Frequency
Preventative Maintenance Tests
Visual inspection When first installed or reinstalled.
NIBP test Pressure check 1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of relevant
module.
3. At least once a year.
Leakage test
CO
2
tests
Leakage test
Performance test
Calibration
Performance Tests
ECG test and
calibration
Performance Test 1. If the user suspects that the measurement is
incorrect.
2. Following any repairs or replacement of relevant
module.
3. At least once every two years. At least once a year
is recommended for NIBP and CO
2
.
Calibration
Resp performance test
SpO
2
test
NIBP test Pressure check
Leakage test
Temp test
IBP test and
calibration
Performance test
Pressure calibration
C.O. test
CO
2
tests and
calibration
Leakage test
Performance test
Calibration
Nurse call function test If the user suspects that the nurse call or analog
output does not work well.
Analog output performance test
Electrical Safety Tests
Electrical Safety Tests At least once every two years.
Other Tests
Power-on test 1. When first installed or reinstalled.
2. Following any repairs or replacement of any main
unit parts.