C-1 
C Default Configurations 
 
This chapter lists some of the most important factory default settings in configuration 
management. You cannot change the factory default configuration itself. However, you can 
make changes to the settings from the factory default configuration and then save the 
changed configuration as a user configuration. The last column of the following tables is for 
your notes and review. 
 
Note: In this chapter, O.M means the monitor’s operating mode. Column C refers to the 
settings that can be changed in configuration management. Column M refers to the settings 
that can be changed in monitoring mode. 
 
 
C.1 Parameters Configuration 
C.1.1 ECG 
ECG Setup 
Item Name 
O.M
General  User Defaults
C M
Lead Set  *  *  Auto   
Alm Source  *  *  HR   
Alarm * * On   
Alm Lev  *  *  Med   
HR/PR 
High 
Adu 
* * 
120  
Ped 160   
Neo 200   
HR/PR 
Low 
Adu 
* * 
50  
Ped 75   
Neo 100   
Sweep * * 25 mm/s   
Beat Vol  *  *  2   
Paced  * No   
Notch Filter  *  *  On   
Gain * * X1   
Filter * * Monitor   
ECG Display  *  *  Normal