C-1
C Default Configurations
This chapter lists some of the most important factory default settings in configuration
management. You cannot change the factory default configuration itself. However, you can
make changes to the settings from the factory default configuration and then save the
changed configuration as a user configuration. The last column of the following tables is for
your notes and review.
Note: In this chapter, O.M means the monitor’s operating mode. Column C refers to the
settings that can be changed in configuration management. Column M refers to the settings
that can be changed in monitoring mode.
C.1 Parameters Configuration
C.1.1 ECG
ECG Setup
Item Name
O.M
General User Defaults
C M
Lead Set * * Auto
Alm Source * * HR
Alarm * * On
Alm Lev * * Med
HR/PR
High
Adu
* *
120
Ped 160
Neo 200
HR/PR
Low
Adu
* *
50
Ped 75
Neo 100
Sweep * * 25 mm/s
Beat Vol * * 2
Paced * No
Notch Filter * * On
Gain * * X1
Filter * * Monitor
ECG Display * * Normal