Loading...
C-2
Item Name
O.M
General User Defaults
C M
Smart Lead Off * * On
ST Analysis
Item Name
O.M General User Defaults
C M
ST Analysis * * Off
Alarm * * Off
Alm Lev * * Med
ST-X High * * when ST Unit is mV:
when ST Unit is mm:
0.20
2.0
ST-X Low * * when ST Unit is mV:
when ST Unit is mm:
-0.20
-2.0
ISO * * -80 ms
J 48 ms
ST J + 60 ms
X represents I, II, III, aVR, aVL, aVF, V, V1, V2, V3, V4, V5 or V6.
Arrh. Analysis
Item Name
O.M
General User Defaults
C M
Arrhythmia Threshold Settings
PVCs/min * *
Adu, Ped:
Neo:
10
10
Tachy * *
Adu:
Ped:
Neo:
120
160
180
Brady * *
Adu:
Ped:
Neo:
50
75
90
Asys. Delay * *
Adu, Ped:
Neo:
5
5
Vtac Rate * *
Adu, Ped:
Neo:
130
160