4-5
WARNING
z Do not discharge a patient before the patient is successfully transferred.
z After a patient is successfully transferred, check if the patient settings (especially
patient category, paced status and alarm limits settings, etc) on the monitor are
appropriate for this patient.
NOTE
z The system automatically switches on the HR alarm and lethal arrhythmia alarm
after transferring the patient data.
Ç‚
z Tranisferring data via USB drive is only avaiable for the monitor configured with a
storage card.
4.4.1 Transferring Data from the Monitor to a USB Drive
1. Select [Main Menu][Patient Setup >>].
2. Select [Transfer to Storage Medium]. In the popup menu, select [Ok].
3. Wait until the following message appears: [Transfer to storage medium successful.
Please remove the USB drive.].
4. Remove the USB drive from patient monitor.
4.4.2 Viewing Data on the Monitor via USB Drive
You can view patient data on the monitor via USB drive without transferring data.
1. Connect the USB drive to the destination monitor.
2. Select [Main Menu][Patient Data >>][History Dat].
You can also select [Patient data in U disk] to view history data if the monitor is configured
with a storage card.

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals