21-3 
21.6 Setting Up Reports 
21.6.1 Setting Up ECG Reports 
You can print out ECG reports only under full-screen, half-screen or 12-lead monitoring 
screen. To set up ECG reports, select [Main Menu]ė[Print Setup >>]ė[ECG
Reports >>]. 
  [Amplitude]: set the amplitude of the ECG waveforms. 
  [Sweep]: set the wave print speed to 25 mm/s or 50 mm/s. 
  [Auto Interval]: If [Auto Interval] is set to [On], the system will automatically adjust 
the space between waveforms to avoid overlapping. 
  [12-Lead Format]: If you select [12
1], 12 waveforms will be printed on a paper from 
top to bottom. If you select [6
2], 12 waveforms will be printed from left to right with 
6 waveforms on each half part and a rhythm waveform will be printed at the bottommost. 
If you select [3
4+1], 12 waveforms will be printed from left to right with 3 
waveforms on each of the 4 columns and a rhythm waveform will be printed at the 
bottommost. 
 
21.6.2 Setting Up Tabular Trends Reports 
To set up tabular trends reports, select [Main Menu]ė[Print Setup >>]ė[Tabular Trends 
Reports >>]. 
  Date: Select the desired date on which a tabular trends report is recorded for printing. 
  Start time: You can set a time period whose trend data will be printed out by setting 
[From] and [Back]. For example, if you set [From] as 2007-4-2 10: 00: 00 and [Back] 
as [2 h], the outputted data will be from 2007-4-2 08: 00: 00 to 2007-4-2 10: 00: 00. In 
addition, the [Back] can be set to either:   
  [Auto]: If [Report Layout] is set to [Time Oriented], the report will be printed by 
time. If [Report Layout] is set to [Parameter Oriented], the report will be printed 
by parameters. 
  [All]: If you select [All], all trend data will be printed out. In this case, it is no need 
to set [From]. 
  [Spacing]: choose the resolution of the tabular trends printed on the report. 
  [NIBP]: prints a tabular trends report when NIBP measurements were acquired. 
Each NIBP measurement will generate a column in the tabular trends review 
window. The values for the other measurements are added to the column to provide 
a complete dataset for the NIBP measurement time. 
  [Auto]: prints a tabular trends report during the period selected and at the interval 
set in the [Interval] option. The [Interval] option is located at the bottom the 
[Tabular Trends] tab of the [Review] window.