B-1 
B EMC and Radio Regulatory Compliance 
 
 
B.1 EMC 
The device meets the requirements of IEC 60601-1-2. 
 
 
NOTE
z  Using accessories, transducers and cables other than those specified may result in 
increased electromagnetic emission or decreased electromagnetic immunity of the 
patient monitoring equipment. 
z  The device or its components should not be used adjacent to or stacked with other 
equipment. If adjacent or stacked use is necessary, the device or its components 
should be observed to verify normal operation in the configuration in which it will 
be used. 
z  The device needs special precautions regarding EMC and needs to be installed and 
put into service according to the EMC information provided below.   
z  Other devices may interfere with this device even though they meet the 
requirements of CISPR. 
z  When the inputted signal is below the minimum amplitude provided in technical 
specifications, erroneous measurements could result. 
z  Portable and mobile communication equipment may affect the performance of this 
monitor.  
z  Other devices that have RF transmitter or source may affect this device (e.g. cell 
phones, PADs, PCs with wireless function).