A-15
Alarm limit Range Step
TB High (low limit + 1) to 43 ć
0.1 ć
TB Low 23 to (high limit - 1) ć
A.7.9 CO
2
(for uMEC12/uMEC15)
Technique Infrared absorption
Standard Meet standard of ISO 80601-2-55
CO
2
Measurement range 0 to 20%
Accuracy*
CO
2
concentration<1%:
1%CO
2
concentration<5%
5%CO
2
concentration<7%
7%CO
2
concentration<12%
12%CO
2
concentration13%
13%<CO
2
concentration20%
20%<CO
2
concentration
±0.1%
±0.2%
±0.3%
±0.4%
±0.5%
±(0.43%+8%rel)
unspecified
±0.3%ABS increase in error for ISO accuracy mode
Accuracy drift Meet the requirement for measurement accuracy within 6 hours
Resolution 1 mmHg
Sample flowrate
Connecting the DRYLINE II watertrap for adult and pediatric
patient: 120 ml/min
Connecting the DRYLINE II watertrap for neonatal patient:
90 ml/min
Connecting the DRYLINE PRIME watertrap: 50 ml/min
Sample flowrate tolerance ±15% or ±15 ml/min, whichever is greater.
Start-up time <90 s
Rise time
Measured with a DRYLINE II neonatal watertrap and a 2.5-meter
neonatal sampling line: <330 ms @ 90 ml/min
Measured with a DRYLINE II adult watertrap and a 2.5-meter
adult sampling line: <300 ms @ 120 ml/min
Measured with a DRYLINE PRIME watertrap and sampling line:
<200 ms @ 50 ml/min
Response time
Measured with a neonatal watertrap and a 2.5-meter neonatal
sampling line: <4.5 s @ 90 ml/min
Measured with an adult watertrap and a 2.5-meter adult sampling
line: <5.5 s @ 120 ml/min
Measured with a DRYLINE PRIME watertrap and sampling line:
<5 s @ 50 ml/min

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz