B-4
Guidance and Declaration - Electromagnetic Immunity
The device is intended for use in the specified electromagnetic environment. The customer or the
user of the device should assure that it is used in such an environment as described below.
Immunity test IEC60601 test
level
Compliance
level
Electromagnetic environment -
guidance
Conduced RF
IEC61000-4-6
3 Vrms
150 kHz to 80
MHz
3Vrms Portable and mobile RF communications
equipment should be used no closer to any
part of the system, including cables, than
the recommended separation distance
calculated from the equation appropriate
for the frequency of the transmitter.
Recommended separation distances:
Pd 2.1
Radiated RF
IEC61000-4-3
3V/m
80MHz to
2.5GHz
3V/m Recommended separation distances:
80 MHz

800 MHz
Pd 2.1
800MHz-2.5GHz
Pd 3.2
Where,
P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and
d is the recommended separation
distance in meters (m).
b
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey
a
,should be less than the
compliance level in each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Note 3: The device that intentionally receives RF electromagnetic energy at the exclusion band
(2395.825MHz-2487.645MHz) is exempt from the ESSENTIAL PERFORMANCE requirements,
but remains safe.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz