Loading...
C-5
C.1.2 RESP
Item Name
O.M
General User Defaults
C M
Alarm * * On
Alm Lev * * Med
Sweep * * 6.25 mm/s
Lead * * II
Gain * * X2
RR High * *
Adu, Ped:
Neo:
30
100
RR Low * *
Adu, Ped:
Neo:
8
30
Apnea Delay * *
Adu, Ped:
Neo:
20
15
Detection
Mode
* * Auto
RR Source * Auto
C.1.3 PR
Item Name
O.M
General User Defaults
C M
Alarm * * On
Alm Lev * * Med
HR/PR
High
Adu
* *
120
Ped 160
Neo 200
HR/PR
Low
Adu
* *
50
Ped 75
Neo 100
PR Source * * SpO
2
Beat Vol * * 2