D-2
Measurement Alarm messages L Cause and solution
Extreme Brady H
R on T M*
Run PVCs L*
PVCs/min M*
Multif.PVC M*
Bigeminy M*
Trigeminy M*
Tachy M*
Brady M*
Vent. Rhythm M*
Multif. PVC M*
Nonsus. Vtac M*
Pause L*
Resp
Resp Apnea H
The respiration signal was so weak that the
monitor cannot perform respiration analysis.
Check the patient’s condition and the Resp
connections.
Resp Artifact H
The patient’s heartbeat has interfered with
his respiration. Check the patient’s condition
and the Resp connections.
SpO
2
SpO
2
Desat H
The SpO2 value has fallen below the
desaturation alarm limit. Check the patient’s
condition and check if the alarm limit
settings are correct.
No Pulse H
The pulse signal was so weak that the
monitor cannot perform pulse analysis.
Check the patient’s condition, SpO
2
sensor
and measurement site.
CO
2
CO
2
Apnea H
The patient stops breathing, or the
respiration signal was so weak that the
monitor cannot perform respiration analysis.
Check the patient’s condition and the RM
connections.

Table of Contents

Other manuals for Mindray uMEC12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray uMEC12 and is the answer not in the manual?

Mindray uMEC12 Specifications

General IconGeneral
TypePatient Monitor
Display12.1-inch color TFT
ECG3-lead
SpO2Yes
NIBPYes
RespirationYes
TemperatureYes
BatteryRechargeable lithium-ion
NetworkingEthernet
ParametersECG, SpO2, NIBP, Respiration, Temperature
ConnectivityEthernet
Battery LifeUp to 4 hours
Power Supply100-240 VAC, 50/60 Hz

Related product manuals